Please wait a minute...

中国生物工程杂志

China Biotechnology
China Biotechnology  2008, Vol. 28 Issue (8): 100-104    DOI:
    
A Novel In-situ Microscope Applied in Biochemical Reactor
Jian Liu Xujun Yuan
Download: HTML   PDF(625KB) HTML
Export: BibTeX | EndNote (RIS)      

Abstract  

The technical requirements of in-situ microscope mounted in biochemical reactor were proposed based on biochemical reactor application. A novel in-situ microscope was designed and manufactured with dark field principle after all principles of in-situ microscopes used currently had been reviewed in detail, the designing principle and structure of a new in-situ microscope was then described. In addition, it was applied to monitor growth of baker yeast and mammalian cell HEK293, respectively. It indecated the cell amount obtained by in-situ microscope was a good agreement with dry cell weight and that determined using haemocytometer,suggested that the designed in-situ microscope could satisfy the requirements of high performance used in bioreactor .



Key wordsIn-situ microscope      Biochemical reactor      Image analysis for cell morphology      Cell countting     
Received: 28 January 2008      Published: 25 August 2008
ZTFLH:  Q786  
Cite this article:

Jian Liu Xujun Yuan . A Novel In-situ Microscope Applied in Biochemical Reactor. China Biotechnology, 2008, 28(8): 100-104.

URL:

https://manu60.magtech.com.cn/biotech/     OR     https://manu60.magtech.com.cn/biotech/Y2008/V28/I8/100