Establishment of Homogeneous Time-resolved Fluorescence (HTRF) Method for Measurement of Monoclonal Antibody Concentration
DENG Chun-pin, QIN Yu-min, DENG Qiao-chun, WANG Bo, PENG Yu-cai
China Biotechnology ›› 2013, Vol. 33 ›› Issue (12) : 64-68.
Establishment of Homogeneous Time-resolved Fluorescence (HTRF) Method for Measurement of Monoclonal Antibody Concentration
A rapid and simple antibody analysis method based on HTRF technology was established to measure concentration of recombinant monoclonal antibody in both cell culture medium and purified antibody solution. APC-labeled antigen,antibody samples or controls, and Eu-labeled anti-human IgG are added into 96 well plate (Black), incubate for some time, followed by fluorescence reading with microplate reader. Standard curve was generated by 4-PL model, antibody concentration in tested samples was calculated accordingly by software. The method qualification results showed that the method is antigen-antibody specific, with accuracy close to 98.9%, precision close to 4.3%, and linearity range 25~1000 ng/ml. This method can be used for cell line screening, bio-process monitoring, and quality control during antibody drug development.
Homogeneous Time-resolved Fluorescence / FRET / Monoclonal antibody {{custom_keyword}} /
[1] 王志明, 高健, 李耿.治疗性单克隆抗体药物的现状及发展趋势.中国生物工程杂志, 2013, 33(6):117-124. Wang Z M, Gao J, Li G. The status and development trend of the therapeutic monoclonal antibody drugs. China Biotechnology, 2013, 33(6):117-124.
[2] 蒋卫, 陈来同, 胡晓倩, 等.紫外分光光度法测定动物血清IgG含量方法的研究.中国生化药物杂志, 2008, 29(2):119-120. Jiang W, Chen L T, Hu X Q, et al. Study on ultraviolet spectrophotometry for determination the content of IgG in animal serum product. Chinese Journal of Biochemical Pharmaceutics, 2008, 29(2): 119-120.
[3] 王淼, 薛巧如, 姚敏菲, 等.测定静脉注射人免疫球蛋白IgG含量的三种方法比较.解放军药学学报, 2010, 26(6):517-520. Wang M, Xue Q R, Yao M F, et al. A comparison of three methods for determination of the content of IgG in human immunoglobulin preparations for intravenous injection(IVIG). Pharm J Chin PLA, 2010, 26(6):517-520.
[4] Isshi K, Kikuchi H, Kojima H, et al. Quantitative determination of serum anti ribosomal-Pprotein antibody by enzyme immunoassay. Rinsho Byori, 1997, 45(12):1177-1181.
[5] Degorce F, Card A, Soh S, et al. HTRF: A technology tailored for drug discovery—a review of theoretical aspects and recent applications. Curr Chem Genomics, 2009, 28(3):22-32.
[6] Gerard M. HTRF technology. J Biomol Screen, 1999, 4(6):309-313.
[7] Clarke E E, Shearman M S. Quantitation of amyloid-beta peptides in biological milieu using a novel homogeneous time-resolved fluorescence (HTRF) assay. J Neurosci Methods, 2000, 102(1):61-68.
[8] 梁菊, 陈攀, 吴文澜, 等.荧光共振能量转移新技术及其在药物筛选中的应用.中国药学杂志, 2009, 44(13):961-965. Lang J, Chen P, Wu W L, et al. The new technology fluorescence resonance energy transfer and its application in drug screening. Chin Pharm J, 2009, 44(13):961-965.
[9] Esohe E I, Joni M C, Clayton C, et al. Development of an antibody screening assay for selection of production cell lines. Bioprocess International, 2008, 4: 20-33.
[10] Selvin PR. Principles and biophysical applications of lanthanide-based probes. Annu Rev Biophys Biomol Struct, 2002, 31:275-302.
[11] Bazin H, Trinquet E, Mathis G. Time resolved amplification of cryptate emission: a versatile technology to trace biomolecular interactions. J Biotechnol, 2002, 82(3):233-250.
[12] Prat O, Lopez E, Mathis G. Europium(Ⅲ) cryptate: a fluorescent label for the detection of DNA hybrids on solid support. Anal Biochem, 1991, 195(2):283-289.
[13] Mathis G. Probing molecular interactions with homogeneous techniques based on rare earth cryptates and fluorescence energy transfer. Clin Chem, 1995, 41(9):1391-1397.
[14] Bazin H, Préaudat M, Trinquet E, et al. Homogeneous time resolved fluorescence resonance energy transfer using rare earth cryptates as a tool for probing molecular interactions in biology. Spectrochim Acta A Mol Biomol Spectrosc, 2001, 57(11):2197-2211.
/
〈 |
|
〉 |